Optimization Of In-field Alpha Spectrometry For Uranium Enrichment Determination In Uranium Hexafluoride

Year
2021
Author(s)
David L. Chichester - Idaho National Laboratory
James T. Johnson - Idaho National Laboratory
Jay D. Hix - Idaho National Laboratory
Scott M. Watson - Idaho National Laboratory
Timothy R. Pope - Pacific Northwest National Laboratory
Glenn A. Fugate - Oak Ridge National Laboratory
File Attachment
a142.pdf1.63 MB
Abstract
In response to needs identified by the International Atomic Energy Agency (IAEA) research is underway to develop In-Field Alpha Spectrometry (IFAS) as a method to allow IAEA safeguards inspectors to collect samples of uranium hexafluoride (UF6) at processing facilities to assess and verify uranium enrichment. For sample collection, the IFAS method uses Single-Use Destructive Assay (SUDA) samplers, which contain thin zeolite coatings that trap UF6 gas and convert it to the safer, more stable form uranyl fluoride (as a dihydrate, UO2F2‧2H2O). For alpha spectrometry, the IFAS instrument employs a large area silicon semiconductor transducer to detect and record alpha particle energy-deposition events. Over the past year optimization work has significantly increased the diameter of useful SUDA samples (from 12.7 mm to 48 mm), improved the manufacturability and reproducibility of SUDA samples, increased the area of the IFAS alpha spectrometer sensor from 1.2 cm to 3.1 cm, and improved source positioning within the IFAS. This paper will report on this optimization work, its impacts on IFAS performance, and future plans for IFAS miniaturization, improvements, and testing.