Year
2000
Abstract
Active well coincidence counter assays have been performed on uranium metal highly enriched in 235U. The data obtained in the present program, together with highly enriched uranium (HEU) metal data obtained in other programs, have been analyzed using two approaches, the standard approach and an alternative approach developed at BNL. Analysis of the data with the standard approach revealed that the form of the relationship between the measured reals and the 235U mass varied, being sometimes linear and sometimes a second-order polynomial. In contrast, application of the BNL algorithm, which takes into consideration the totals, consistently yielded linear relationships between the totals-corrected reals and the 235U mass. The constants in these linear relationships varied with geometric configuration and level of enrichment. This indicates that, when the BNL algorithm is used, calibration curves can be established with fewer data points and with more certainty than if a standard algorithm is used. However, this potential advantage has only been established for assays of HEU metal. In addition, the method is sensitive to the stability of natural background in the measurement facility.