MONOCHROMATIC WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE PROVIDING SENSITIVE AND SELECTIVE DETECTION OF URANIUM

Year
2010
Author(s)
Michael Collins - Los Alamos National Laboratory
Joseph Montoya - Los Alamos National Laboratory
George J. Havrilla - Los Alamos National Laboratory
Zewu Chen - X-ray Optical Systems
Fuzhong Wei - X-ray Optical Systems
Abstract
Monochromatic wavelength dispersive X-ray fluorescence (MWDXRF) is a sensitive and selective method for elemental compositional analyses.[1-3] The basis for this instrumental advance is the doubly curved crystal (DCC) optic. Previous work has demonstrated the feasibility of sensitive trace element detection for yttrium as a surrogate for curium in aqueous solutions.[1] Additional measurements have demonstrated similar sensitivity in several different matrix environments which attests to the selectivity of the DCC optic as well as the capabilities of the MWDXRF concept. The objective of this effort is to develop an improved Pu characterization method for nuclear fuel reprocessing plants. The MWDXRF prototype instrument is the second step in a multi-year effort to achieve an improved Pu assay. This work will describe a prototype MWDXRF instrument designed for uranium detection and characterization. The prototype consists of an X-ray tube with a rhodium anode and a DCC excitation optic incorporated into the source. The DCC optic passes the RhKa line at 20.214 keV for monochromatic excitation of the sample. The source is capable of 50 W power at 50 kV and 1.0 mA operation. The x-ray emission from the sample is collected by a DCC optic set at the ULa line of 13.613 keV. The collection optic transmits the ULa x-rays to the silicon drift detector. The x-ray source, sample, collection optic and detector are all mounted on motion controlled stages for the critical alignment of these components. The sensitivity and selectivity of the instrument is obtained through the monochromatic excitation and the monochromatic detection. The prototype instrument performance has a demonstrated for sensitivity for uranium detection of around 2 ppm at the current state of development. Further improvement in sensitivity is expected with more detailed alignment.